Recall Z-0611-03 ยท DPZ Oximeter, Ear

Datex-Ohmeda

Datex-Ohmeda Cardiocap/5 Anesthesia Monitor with N-XOSAT SpO2 measurement option

DateMar 5, 2003 (initiated Sep 16, 2002)
StatusTerminated
ReasonFalsely high blood oxygen saturation readings in some circumstances which could result in no alarm in a critical life-threatening situation.
Root causeOther
Action'Medical Device Advisory Notice' letters dated September 16, 2002 were sent to the consignees, gave recommendations about alternative monitoring for certain patients, and stated that there would be a field correction to reduce the possibility of the problem.
Quantity85
DistributionNationwide in the United States and worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012837MODIFICATION TO: DATEX-OHMEDA CARDIOCAP 5 (REV B) AND ACCESSORIES
Datex-Ohmeda, Sep 14, 2001