Philips Medical Systems, Inc. Cardiac & Monitoring Systems
Viridia/Agilent Information Center Model : M3151A#CO1
| Date | Feb 28, 2003 (initiated Jan 2, 2003) |
| Status | Terminated |
| Reason | Central Station Monitoring Information Centers may experience sreeen freeze, i.e. loss of display, keyboard or mouse lock up |
| Root cause | Other |
| Action | Philips Medical Systems notified users by Certified Mail on/about 2/19/03 that the firm was initiating a device field correction (recall) to conduct an upgrade to units witha HP VL400 or VL420 computer which is cconfigured for dual display by replacing the video and network communication card. Previous to this notification, Philips sent letters on 1/2/03 addressed to Biomedical Engineering Manager to advise users to reboot the system in the event of a system freeze. |
| Quantity | 82 units |
| Distribution | Nationwide Canada, Europe, Asia, Latin America |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K011093 | AGILENT INFORMATION CENTER SOFTWARE FOR M3150,M3151,M3153 AND M3154 Agilent Technologies, Inc., May 1, 2001 |