Recall Z-0586-03 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified
Philips Medical Systems
MultiDiagnost 4 - x-ray system, multi functional/universal system for general use.
| Date | Feb 27, 2003 (initiated Jan 10, 2003) |
| Status | Terminated |
| Reason | INTERNAL SHORT CIRCUIT - MAY ALLOW FOR STAND AND TABLE TO MOVE SPONTANEOUSLY IN UPWARD OR DOWNWARD MOVEMENT |
| Root cause | Other |
| Action | The firm issued a letter dated January 10, 2003 to their customers advising of the problem and that a respresentative would be contacting customers to arrange a time to visit to install new interface board. This is being conducted under FCO 70800007 |
| Quantity | 22 units |
| Distribution | The firm distributed to 22 hospitals/medical centers located in CA, FL, IA, IL, IN, KY, MN, MO, NC, ND, NY, OH, SC, TN, TX, VA, and WV. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K961374 | PHILIPS MULTIDIAGNOST 4 Philips Medical Systems, Inc., Aug 19, 1996 |