Recall Z-0586-03 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips Medical Systems

MultiDiagnost 4 - x-ray system, multi functional/universal system for general use.

DateFeb 27, 2003 (initiated Jan 10, 2003)
StatusTerminated
ReasonINTERNAL SHORT CIRCUIT - MAY ALLOW FOR STAND AND TABLE TO MOVE SPONTANEOUSLY IN UPWARD OR DOWNWARD MOVEMENT
Root causeOther
ActionThe firm issued a letter dated January 10, 2003 to their customers advising of the problem and that a respresentative would be contacting customers to arrange a time to visit to install new interface board. This is being conducted under FCO 70800007
Quantity22 units
DistributionThe firm distributed to 22 hospitals/medical centers located in CA, FL, IA, IL, IN, KY, MN, MO, NC, ND, NY, OH, SC, TN, TX, VA, and WV.