Recall Z-0542-03 ยท MKJ Automated External Defibrillators (Non-Wearable)

Medtronic Physio Control Corp

LIFEPAK 12 defibrillator/monitor is a portable, battery operated, complete acute cardiac care response system with both manual and semi-automatic defibrillation operation. It has the ability to externally pace, manually defibrillate, or automatically analyze and if necessary, defibrillate a patient through therapy cables with disposable electrodes, standard paddles, or internal handle cables. P

DateFeb 5, 2003 (initiated Jan 27, 2003)
StatusTerminated
ReasonImpact and damage to the high voltage connectors results in poor or no electrical contact from the cable to the LIFEPAK 12 unit.
Root causeOther
ActionConsignees were notified of recall action by letter dated January 2003. Letter advised consignees to follow recommendations for periodic inspections and to test device to verify readiness. A few weeks later the consignees were sent a correction letter along with user installable shields.
Quantitydomestic: 30,575 foreign: 15,467
DistributionWorldwide distribution, including government/military accounts.
FDA recordOpen on fda.gov