Recall Z-0565-03 ยท FKX System, Peritoneal, Automatic Delivery

Baxter Healthcare Corp.

HomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., Made in U.S.A.; catalog numbers 5C8310, 5C8310R, R5C8320, R5C8320R, T5C8300, T5C8300R

DateFeb 21, 2003 (initiated Dec 23, 2002)
StatusTerminated
ReasonPotential for low ultrafiltrate volumes when using the Low Fill Mode during Continuous Cycling Peritoneal Dialysis (CCPD) Therapy
Root causeOther
ActionSafety Alert letters dated 12/23/02 were mailed to the medical directos of facilities identified as having patients who use the low fill volume cassette with the HomeChoice machines. The letters informed them of the potential low ultrafiltrate volumes and requested that they review the patient list provided with the letter and verify that all of their patients who use the device in the low fill mode are included.
DistributionNationwide, Guatemala, Chile, Brazil, Turkey, Singapore, Taiwan, Canada, Belgium, Czech Republic, Israel, Italy, Jordan, Lebanon, Malta, Spain, Sweden, Turkey and the United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012988HOMECHOICE PERSONAL CYCLER PERITONEAL DIALYSIS SYSTEM, MODELS 5C8310, 5C8302, 5C4471, 5C4469
Baxter Healthcare Corp, Dec 5, 2001