Recall Z-0438-03 ยท

Roche Diagnostics Corp.

PO2 electrode for Roche AVL OMNI Analyzers. Catalog number BP1734.

DateJan 23, 2003 (initiated Dec 16, 2002)
StatusTerminated
ReasonAir bubbles trapped in the PO2 electrode will cause falsely elevated patient test results.
Root causeOther
ActionAn urgent device correction letter dated 12/16/02 was sent to each customer. The letter advised of the problem and instructed customers to follow a strict quality control regime to ensure proper electrode performance. Customers were instructed how to automatically lock the affected parameter so that patient results will not be reported if QC performance violates the customer''s programmed rules.
Quantity695
DistributionUnited States
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013373OMNI C ANALYZER
Roche Diagnostics Corp., Nov 2, 2001
K990092AVL OMNI MODULAR ANALYZER
Avl Scientific Corp., Jan 29, 1999