Recall Z-0511-03 ยท FPA Set, Administration, Intravascular

Baxter Healthcare Corp.

Baxter Auto Syringe Micro-Volume Extension Set with 0.22 Micron Filter, catalog 1M8516 and 1M8516Y; Baxter Healthcare Corporation, Auto Syringe Division, Hookset, NH 03106

DateFeb 6, 2003 (initiated Oct 17, 2002)
StatusTerminated
ReasonThe I.V. filter sets contain a 1.2 micron filter instead of the labeled 0.22 micron filter.
Root causeOther
ActionRecalled by letter dated 10/17/02. Accounts were informed of the potential for the filters to be 1.2 micron filters instead of the 0.22. micron filters as labeled. The accounts were requested to examine their inventories for the affected catalog numbers and destroy the units found for credit.
Quantity16,860 units
DistributionNationwide, Australia, Puerto Rico, Brazil, Canada, Sweden, New Zealand, Japan, Mexico, American Samoa, El Salvador, Turkey, Saudi Arabia, Austria, Cayman Islands, Macedonia, Honduras, Kuwait, Hong Kong and Nigeria.