Recall Z-0605-03 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Corp.

Baxter Meridian Hemodialysis Instrument, product code 5M5576 and 5M5576R; Baxter Healthcare Corporation, McGaw Park, IL 60085 The instrument is part of a high permeability hemodialysis system which consists of a controlled dialysate delivery system that incorporates an ultrafiltration controller to prevent excessive loss of water from the patient''s blood, an extracorporeal blood set, and a high

DateMar 5, 2003 (initiated Nov 15, 2002)
StatusTerminated
ReasonMeridian hemodialysis instrument may reset when in shutdown mode causing the system to reset with potential for patient increased bioburden exposure.
Root causeOther
ActionBaxter issued a Safety Alert letter dated 11/15/02 to the direct accounts informing them of the potential for incomplete disinfection of the Meridian instrument during the Renalin/Formalin disinfection cycle due to electomagnetic interference (EMI). The accounts were informed that they will be receiving a radio frequency interference (RFI) filter kit and installation instructions to eliminate EMI by 2/28/03. The accounts were given instructions to follow to ensure that the Renalin/Formalin disinfection cycle has been completed prior to use of the machine.
Quantity2864 units
DistributionNationwide, Mexico, China, South Korea, Hong Kong
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.