Recall Z-0421-03 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

AxSYM System, Lift Mechanisms: List Numbers 7A83-01, 7A83-95, 7A83-98, and 7A83-03

DateJan 11, 2003 (initiated Dec 18, 2002)
StatusTerminated
ReasonGas Springs supporting processing center cover lose pressure after repeated usage causing the cover to unexpectedly close.
Root causeOther
ActionAbbott Laboratories sent a December 18, 2002 letter to all consignees informing consignees of their plans to install a locking mechanism on all AxSYM Systems. The installations will be performed by Abbott Representatives. A customer reply form was included in the device correction letter.
Quantity18,832 units
DistributionNationwide and to the following foreign countries: Mexico, Guatemala, Venezuela, Columbia, Ecuador, Peru, Chile, Uruguay, Argentina, Brazil, Dominican Republic, El Salvador, Canada, Germany, Ireland, England, Belgium, France, Austria, Italy, Spain, Russia, Japan, Singapore, Hong Kong, Taiwan, Thaila
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974651ABBOTT AXSYM SYSTEM
Abbott Laboratories, Feb 12, 1998