Recall Z-0355-2027 ยท NGZ Susceptibility Test Plate, Antifungal

Remel, Inc

Yeastone Broth, 11ML, 10/BOX, used in antibiotic susceptibility testing of mycological clinical specimens

DateNov 3, 2026 (initiated Oct 19, 2026)
StatusOpen, Classified
ReasonAffected lot did not meet specifications. Product may report false susceptibility to antifungal agents, specifically within the Azole family.
Root causeUnder Investigation by firm
ActionThe firm began notifying affected customers on October 19, 2026. Customers were instructed to notify any personnel who needed to be aware of the issue. Customers should review previously reported susceptibility results generated using the affected lot and assess whether retesting is appropriate. Customers should destroy any remaining inventory of the affected lot. Do not use it for any clinical laboratory testing. Customers should complete and return the acknowledgement form accompanying the letter. Firm will replace product or issue credit. If you have any questions, please contact our
Quantity33 US, 54 OUS
DistributionWorldwide - US Nationwide distribution in the states of MA and CA. The countries of India, Saudi Arabia, UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090968SENSITITRE YEASTONE SUSCEPTIBILITY PLATES
Trek Diagnostic Systems, Inc., Jun 4, 2009