ClearanceBrief
Recall Z-0288-2027 ยท IZL System, X-Ray, Mobile

Philips North America

Model Number: 712010.

DateOct 26, 2026 (initiated Sep 18, 2026)
StatusOpen, Classified
ReasonPotential for the device to fail immediately after an exposure, causing the system to remain stationary.
Root causeSoftware design
ActionPhilips Medical Systems notified consignees on about 09/18/2026 via "URGENT Medical Device Correction" letter. Consignees were instructed that they may continue to use the affected system in accordance with its intended use, however if the no-drive issue occurs and prevents repositioning of the system, system movement may be restored through one or more of the following options: - Option 1: System reboot: Power OFF the unit, wait at least 10 seconds, and then power system ON While system startup may take up to 4 minutes, drive functionality will be available within approximately 8 seconds a
Quantity563 units
DistributionWorldwide Distribution: (US Nationwide) and International distribution to Angola, Australia, Austria, Canada, Chile, Cook Islands, Finland, France, Germany, Ghana, Gibraltar, Guadeloupe, Honduras, Israel, Italy, Latvia, Lithuania, Macedonia, Namibia, Netherlands, New Zealand, Poland, Portugal, Roman
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K233678Radiography 7000 M
Philips Medical Systems Dmc GmbH, Feb 14, 2024