ClearanceBrief
Recall Z-0333-2027 ยท GDC Table, Operating-Room, Electrical

Baxter Healthcare Corporation

Mobile column TruSystem 7500 U, Product Code 1730720

DateNov 3, 2026 (initiated Sep 14, 2026)
StatusOpen, Classified
ReasonThe main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
Root causeEquipment maintenance
ActionCustomers were notified by telephone beginning 9/14/26. Baxter Healthcare Corporation is issuing a voluntary correction for the TruSystem Surgical Tables after identifying that four units underwent spindle replacement without the required belt tension verification. The affected units were released without confirmation that belt tension met the specified requirements, which could lead to unintended downward movement of the table column. Unintended downward movement of the surgical table could occur, potentially resulting in disruption of patient positioning, reduced perfusion or oxygenation, m
Quantity1 unit
DistributionUS Nationwide distribution in the states of IL, MO, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.