Recall Z-0347-2027 ยท OWB Interventional Fluoroscopic X-Ray System

SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION

Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),

DateNov 3, 2026 (initiated Sep 30, 2025)
StatusOpen, Classified
ReasonThe mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.
Root causeProcess control
ActionOn September 30, 2026 Shimadzu Corp. issued a Urgent Medical Device Recall Notification to affected consignees via email. Shimadzu asked consignees to take the following actions; 1. If the bolts become loose, an abnormal noise will occur when moving the C-arm. Please check for abnormal noises during pre-study inspections. 2. If an abnormal noise is heard, please stop using the product and contact our service department. 3. Please complete the Medical Device Recall Customer Acknowledgement.
Quantity118 unts.
DistributionUS Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203535Trinias
Shimadzu Corporation Medical Systems, Apr 28, 2021
K221922Trinias
Shimadzu Corporation Medical Systems Division, Jul 28, 2022