Recall Z-0359-2027 ยท NSU Instrumentation For Clinical Multiplex Test Systems

Luminex Corporation

LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303

DateNov 5, 2026 (initiated Oct 5, 2026)
StatusOpen, Classified
ReasonInstructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection.
Root causeError in labeling
ActionOn October 5, 2026, the firm began notifying customers via "Urgent Field Safety Notice - Medical Device Recall" letters. Consignees were informed that the current Instructions for Use for the Gram-Negative Blood Culture Assay do not adequately describe interpretation of certain positive family-level results. Specifically, when a positive family-level only result is obtained ("family-level only result" refers to a positive organism family-level detection without an accompanying genus- or species-level identification), repeat testing with the LIAISON PLEX Gram-Negative Blood Culture Assay is re
Quantity30,168 test cartridges (2,514 Kits)
DistributionUS distribution nationwide. No OUS distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K243013LIAISON PLEX Gram-Negative Blood Culture Assay
Luminex Corporation, Apr 18, 2025