Recall Z-0286-2027 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

ZIEHM-ORTHOSCAN INC

Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020

DateOct 23, 2026 (initiated Sep 3, 2026)
StatusOpen, Classified
ReasonThe accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.
Root causeProcess control
ActionOn September 11, 2026, Ziehm-Orthoscan issued a Urgent: Medical Device Recall Notification via Email. Ziehm-Orthoscan ask consignees to take the following actions: 1. The device may continue to be used until the corrective action has been completed, provided the user follows the precautions described in this notification and no signs of instability or unintended movement are observed. 2. Users should exercise caution when positioning, moving, or operating the system and monitor assembly, avoid fully extending the display arm until the corrective action has been performed. 3. If unintended m
Quantity55 units
DistributionWorldwide - US Nationwide distribution in the states of AZ , CA , CO , CT , FL , GA , IL , IN , KS, MA , MD , MI , MN , NC , NE , NV , NY , OH , OR , SC, TX , UT , WA and the countries of Germany , Australia , Spain , Switzerland, Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K243452Orthoscan VERSA Mini C-Arm
Ziehm-Orthoscan, Inc., Jan 14, 2025