Recall Z-0283-2027 ยท LRO General Surgery Tray

AVID Medical, Inc.

Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH30.

DateOct 22, 2026 (initiated Sep 8, 2026)
StatusOpen, Classified
ReasonA non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.
Root causeProcess control
ActionAVID Medical notified consignees on about 09/08/2026 via emailed "URGENT: MEDICAL DEVICE CORRECTION" letter. Consignees were instructed to identify, segregate, and quarantine all impacted kits identified, and to not use the impacted component located in the piggybacked portion of the convenience kit. Customers were also instructed to add warning labels to affected kits to ensure all users are aware of the actions to remove and discard the impacted component prior to use. distributors were instructed to notify consignees if affected units were further distributed. Additionally, consignees were
Quantity80 kits
DistributionUS Nationwide distributor in the state of Arizona.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.