Recall Z-0298-2027 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips Azurion 7 M12. Model Numbers: 722078, 722223.

DateOct 27, 2026 (initiated Oct 6, 2026)
StatusOpen, Classified
ReasonSolid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Root causeEquipment maintenance
ActionPhilips Medical notified consignees on about 10/06/2026 via certified mailed letter titled "URGENT Medical Device Correction." Consignees were instructed that affected systems may continue to be used in accordance with their Instructions for Use (IFU) and the additional instructions provided, which include: if motorized movements become unavailable, the stand can be manually repositioned using the handgrips and brake controls located on both sides of the stand. For ceiling-mounted systems, manual longitudinal (1), transversal (2), and rotational (3) stand movements remain available. Note: Tra
Quantity602 units
DistributionWorldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172822Azurion series R1.2
Philips Medical Systems Nederland B.V., Nov 22, 2017
K181830Azurion R2.0
Philips Medical Systems Nederland B.V., Aug 2, 2018
K200917Azurion R2.1
Philips Medical Systems Nederland B.V., May 1, 2020