ClearanceBrief
Recall Z-3272-2026 ยท JHI Visual, Pregnancy Hcg, Prescription Use

Alere San Diego, Inc.

Fisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman Coulter Inc's ICON 25 hCG, Model/ Part: FHC-202/43025; Henry Schein One Step+ hCG Combo test in Serum or Urine, Model/ Part: FHC-202/9004075; Alere hCG Combo Cassette (25mIU/mL) in Urine (CLIA Waived) or Serum, Model/ Part: FHC-202/92225; Pro Advantage By NDC Urine/Serum Pregnancy Cassette Dev

DateSep 25, 2026 (initiated Aug 20, 2026)
StatusOpen, Classified
ReasonhCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Root causeDevice Design
ActionOn 8/20/2026, correction notices started being emailed to customers and distributors informing them to take the following actions: 1) In situations where an unexpected test result is obtained, consider all limitations, including peri-menopause that may result in low levels of hCG. A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated, and an alternate confirmatory test that supports a timely clinical decision has been considered. 2) If you have forwarded affected products to other laboratories within your distribution
Quantity5,331,585
DistributionWorldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Irel
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993065ACON COMBO PREGNANCY
ACON Laboratories, Inc., Oct 8, 1999