Recall Z-0147-2027 ยท OHO First Aid Kit Without Drug

INTEGRATED MEDCRAFT LLC

FORWARD Gunshot Wound Kit; Model: AFT-GSW;

DateOct 9, 2026 (initiated Sep 25, 2026)
StatusOpen, Classified
ReasonSubrecall of a kitted device from SafeGuard Medical due to internal testing identified an issue affecting the device's sterile pad, potentially causing the sterile pad to degrade over time. The degradation may cause remnants of the nonwoven pad material to dislodge on the wound or in close proximity to the wound.
Root causeComponent change control
ActionOn September 25, 2026 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: 1. Immediately discontinue use of any product matching the affected item(s) and lot number(s) listed in the item table above. Inspect inventory for affected product. If you have no products matching the affected item(s), please fill out the acknowledge receipt form (referenced below) and indicate that you have none of these products in inventory. 2. A copy of the Urgent Medical Device Recall from Safeguard Medical has been included for reference. 3. If you have product affected by this not
Quantity1033
DistributionUS Nationwide distribution in the states of GA, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.