Recall Z-0242-2027 ยท KRA Catheter, Continuous Flush

Boston Scientific Cork, Ltd.

Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/105; Material Number: M001195430

DateOct 22, 2026 (initiated Sep 16, 2026)
StatusOpen, Classified
ReasonThe product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Root causeProcess control
ActionOn September 16, 2026, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. Actions to be taken: 1. Immediately stop further use or distribution and segregate affected product (see Attachment 1 for identification of the affected devices). 2. Forward this notice to relevant personnel and any facilities to which the product was transferred. 3. Continue to follow the standard of practice when monitoring patients treated with the product. 4. Complete and return the enclosed Reply Verification Tracking Form per the included instructions. 5. Return product to
Quantity76
DistributionWorldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Jap
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163701Direxion and Direxion HI-FLO Torqueable Microcatheters
Boston Scientific Corporation, May 16, 2017