Recall Z-0199-2027 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips North America

Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;

DateOct 16, 2026 (initiated Sep 16, 2026)
StatusOpen, Classified
ReasonUnreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC iX).
Root causeSoftware design
ActionOn september 16, 2026, URGENT Medical Device Correction, letters were sent to customers. Actions to be taken: 1. Direct clinical attendance of the patient during transport, when this issue is most likely to manifest, should be considered if not currently a part of the clinical workflow. In the event a persistent No Central Monit technical alarm is noted at the wireless device or a No Data technical alarm is noted at the central monitor; a restart of the wireless monitor may restore central monitoring when the patient is in known good wireless coverage area. 2. Pass this notice to all those
Quantity174
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.