ClearanceBrief
Recall Z-0179-2027 ยท OJH Orthopedic Tray

Medline Industries, LP

Medline surgical convenience kits labeled as: 1) SPINAL CDS, Medline Product Number/SKU CDS940087AM; 2) CUSTOM CRANI, Medline Product Number/SKU DYNJ907697C; 3) FEM POP PACK-LF, Medline Product Number/SKU DYNJ0159820F; 4) AV FISTULA PACK, Medline Product Number/SKU DYNJ67139D; 5) CRANIOTOMY CDS, Medline Product Number/SKU CDS780025AD; 6) LAMINECTOMY PACK, Medline Product Number/SKU DYNJ1733

DateOct 15, 2026 (initiated Sep 2, 2026)
StatusOpen, Classified
ReasonMedical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Root causeUnder Investigation by firm
ActionMedline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 9/2/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory,
Quantity6069 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.