Recall Z-0173-2027 ยท OGR Ear, Nose, And Throat Surgical Tray

Medline Industries, LP

Medline surgical convenience kits labeled as: 1) EAR, Medline product number/SKU DYNJ909020B; 2) EAR, Medline product number/SKU DYNJ909020A; 3) CAROTID PACK, Medline product number/SKU DYNJ42922C; 4) EAR, Medline product number/SKU DYNJ909020; 5) FEM CAROTID PACK, Medline product number/SKU DYNJ65706; 6) CAROTID CDS-LF, Medline product number/SKU CDS983139B.

DateOct 15, 2026 (initiated Sep 2, 2026)
StatusOpen, Classified
ReasonMedical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Root causeUnder Investigation by firm
ActionMedline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 9/2/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory,
Quantity620 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall