Recall Z-0171-2027 ยท OFF Central Venous Catheter Tray
Medline Industries, LP
Medline surgical convenience kits labeled as: 1) VEIN HARVESTING PACK-LF, Medline product number/SKU DYNJ0826732J; 2) NBMC VASCULAR PACK-LF, Medline product number/SKU DYNJ0665837G; 3) AV FISTULA, Medline product number/SKU DYNJ900454C; 4) VEIN PACK SHD, Medline product number/SKU DYNJ68420D; 5) PACK VASCULAR BYPASS, Medline product number/SKU DYNJ905692J; 6) PRODUCT VASCULAR LAFAYETTE, Med
| Date | Oct 15, 2026 (initiated Sep 2, 2026) |
| Status | Open, Classified |
| Reason | Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use. |
| Root cause | Under Investigation by firm |
| Action | Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 9/2/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, |
| Quantity | 8374 kits |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.