Recall Z-0192-2027 ยท OLO Orthopedic Stereotaxic Instrument

Think Surgical, Inc.

Think Surgical Square Driver REF: 109126

DateOct 16, 2026 (initiated Aug 20, 2026)
StatusOpen, Classified
ReasonRobotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.
Root causeProcess control
ActionOn August 25, 2026, THINK Surgical Inc. issued a Urgent Medical Device Recall Notification to affected consignees via email. Think Surgical sent an updated letter via email on September 22, 2026 to provide additional product information. THINK Surgical ask consignees to take the following actions: 1. Examine your inventory of square drivers for the lot number 25356. Quarantined return any inventory of that lot to THINK Surgical, Inc., 2. If you have distributed or transferred affected devices to other facilities or end users, please promptly notify them and provide a copy of this notificat
Quantity15 units
DistributionUS Nationwide distribution in the states of NY and DE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K261024TMINI Miniature Robotic System
THINK Surgical, Inc., Jul 24, 2026