Recall Z-3155-2026 ยท JAK System, X-Ray, Tomography, Computed

Philips North America Llc

Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

DateSep 29, 2026 (initiated Aug 18, 2026)
StatusOpen, Classified
ReasonThe Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
Root causeRadiation Control for Health and Safety Act
ActionPhilips sent an Electronic Product Radiation Correction notification dated August 18, 2026 to the users of the affected systems to alert them of the issue and provide the actions that should be taken in order to prevent risks to patients. Philips will schedule an appointment with customers to install a software solution free of charge.
Quantity62
DistributionU.S. and U.S. Territories.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K244008Spectral CT
Philips Medical Systems Technologies , Ltd., May 16, 2025