Recall Z-0145-2027 ยท FPO Stretcher, Wheeled
Ferno-Washington Inc
Power F2 Powered Fastening System, Power F2, Item No. 0567093 used with Power X2 Stretchers, Power X2, Item No. 0015816 and PowerX2 Premium, Item No. 0015821
| Date | Oct 9, 2026 (initiated Apr 28, 2026) |
| Status | Open, Classified |
| Reason | Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or exhibiting reduced powered performance. Issue may result in interruption or delay in patient transport. |
| Root cause | Software design |
| Action | Firm began notifying customers on April 28, 2026. Customers were informed that their Power F2 system would be receiving a firmware update. PX2 firmware was also updated to ensure consistent communication with PF2 Connect. Instruction to Customers: Until your Power F2 system has been updated: - continue to use the device in accordance with the instructions for use - if powered functionality is not available, transition to manual operation as needed -ensure personnel are familiar with manual operations procedures |
| Quantity | 319 |
| Distribution | Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, ID, IL, KY, LA, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Netherlands and Poland. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.