Recall Z-0002-2027 ยท LCO Platelet Factor 4 Radioimmunoassay

Instrumentation Laboratory

HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)

DateOct 10, 2026 (initiated Sep 3, 2026)
StatusOpen, Classified
ReasonPotential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Root causeUnder Investigation by firm
ActionInstrumentation Laboratory notified consignees on about 09/03/2026 via emailed and mailed URGENT MEDICAL DEVICE REALL letter. Consignees were made aware of the issue and hazard involved. They were instructed to check inventory for affected lots, destroy any remaining product following each facility's standard procedures, and document the destruction on the enclosed Urgent Mandatory Response Tracking Form, including quantity destroyed and date of destruction. Consignees were asked to identify an alternative method for HIT antibody testing as a supply interruption is expected and work local We
Quantity1,414 units
DistributionWorldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Kore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170854HemosIL AcuStar HIT-IgG(PF4-H), HemosIL AcuStar HIT Controls
Instrumentation Laboratory CO, Sep 8, 2017