ClearanceBrief
Recall Z-0132-2027 ยท CBK Ventilator, Continuous, Facility Use

Ventec Life Systems, Inc.

V+ Pro Emergency, Model Number: PRT-01201-000, Package PN:PRT-01173-000,PRT-01222-000. used for ventilatory support.

DateOct 20, 2026 (initiated Sep 17, 2026)
StatusOpen, Classified
ReasonVentilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Root causeUnder Investigation by firm
ActionOn September 17, 2026, Ventec Life Systems Inc. dba React Health issued a Urgent: Medical Device Recall Notification via certified mail. Ventec ask consignees to take the following actions: 1.Review the customer communication and follow the applicable Actions to be Taken section. 2. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to React Health via email, mail or use the online customer response form within 7 days. 3. If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify
Quantity30,120 units
DistributionDistribution US nationwide including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162877VOCSN Unified Respiratory System
Ventec Life Systems, Inc., Apr 7, 2017