ClearanceBrief
Recall Z-3309-2026 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips North America

Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions: R11, R12 and R13;

DateSep 25, 2026 (initiated Aug 28, 2026)
StatusOpen, Classified
ReasonWhen exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Root causeSoftware design
ActionOn August 28, 2026, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. To prevent the issue from occurring while Philips is developing the software solution, use the Classic DICOM transfer option for exporting DTI scans. 2. You may continue to use your system(s) in accordance with the intended use. 3. Circulate this notice to all users of this device so that they are aware of the issue and associated hazard/harm. 4. Please retain this Urgent Medical Device Correction letter with your system(s) until the software upgrade is installed; ensure the notice is i
Quantity243
DistributionWorldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Gree
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR Systems
Philips Medical Systems Nederland B.V., Feb 14, 2019
K251808Achieva; Intera; Ingenia 1.5T; Ingenia 3.0T; Ingenia 1.5T CX; Ingenia 3.0T CX; Ingenia Elition S; Ingenia Elition X; Ing
Philips Medical Systems Nederland B.V., Jul 11, 2025