ClearanceBrief
Recall Z-3227-2026 ยท LLI Bed, Therapeutic, Ac-Powered, Adjustable Home-Use

Medline Industries, LP

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED

DateOct 1, 2026 (initiated Aug 14, 2026)
StatusOpen, Classified
ReasonPotential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
Root causeDevice Design
ActionFirm initially notified customers November 26, 2025. On August 14, 2026, the firm sent a follow-up letter which provides additional product SKU - Item numbers for the electrical system safety risks and additional important information related to the previous notices. The notice explained the issue, potential risks, and requested the user ONLY USE MEDLINE SIDE RAILS and MEDLINE ACCESSORIES that are intended for use with Medline Homecare Beds. When using Medline Side Rails and Medline Accessories, follow all applicable instructions for use. Distributors were directed to notify their customers.
Quantity771,127 total
DistributionDistribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.