Recall Z-0119-2027 · OLO Orthopedic Stereotaxic Instrument

Intellijoint Surgical, Inc.

Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP, Portable System, REF: 000-0249

DateOct 7, 2026 (initiated Aug 28, 2026)
StatusOpen, Classified
ReasonA navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.
Root causeNonconforming Material/Component
ActionOn 8/28/2026, recall notices were emailed to customers who were asked to do the following: - If affected serial numbers are found, stop use and contact firm's customer service: at customerservice@intellijointsurgical.com and copy regulatory@intellijointsurgical.com - If the observed serial number does not match the affected serial number please record the serial number identified and inform customerservice@intellijointsurgical.com and cc regulatory@intellijointsurgical.com of the correct serial number for that facility.
Quantity8
DistributionWorldwide - US Nationwide distribution in the states of CA, TX, MS, NJ and the countries of Singapore, Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191507Intellijoint® Navigation System
Intellijoint Surgical, Inc., Oct 1, 2019