ClearanceBrief
Recall Z-0104-2027 ยท LRO General Surgery Tray

AVID Medical, Inc.

MAJOR PACK; Model/Item Number (Kit): NIHB0018-11;

DateOct 7, 2026 (initiated Aug 10, 2026)
StatusOpen, Classified
ReasonProduct may be potentially contaminated with a lubricant used during the manufacturing process.
Root causeProcess change control
ActionOn August 10, 2026, URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: 1. Identify, segregate, and quarantine all Avid kits identified in Attachment 1. 2. Add recall labels to affected Avid kits to ensure all users are aware of the actions to discard the impacted component. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. This label should be applied in a prominently visible location to end-users. The location chosen sho
Quantity40
DistributionUS Nationwide distribution in the states of AZ, CA, FL, MD, NE, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.