Recall Z-0108-2027 ยท OWB Interventional Fluoroscopic X-Ray System

Canon Medical System, USA, INC.

Alphenix INFX-8000C/B5

DateOct 7, 2026 (initiated Aug 26, 2026)
StatusOpen, Classified
ReasonLoosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error
Root causeProcess control
ActionOn August 26, 2026, Canon Medical Systems USA, Inc. issued a Urgent: Medical Device Correction notification to affected consignees via mail. Canon ask consignees to take the following actions: 1. Should this problem occur before inspection of the tightening of the bolt(s) is implemented on your system, please stop using the device and call your service representative. 2. Additionally, it is strongly requested that you share the contents of this letter with all users as well as clinical engineering or biomedical group at your facility. 3. A representative of Canon Medical Systems USA will c
Quantity1 unit
DistributionUS Nationwide distribution in the state of ID.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K210900Alphenix, INFX-8000C/B, INFX-8000C/S, V9.2
Canon Medical Systems Corporation, Apr 12, 2021