ClearanceBrief
Recall Z-0083-2027 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter Inc.

DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC 500i, Part No. C63521 3. Access 2 Module, DxC 500i, Part No. C13252

DateOct 2, 2026 (initiated Sep 2, 2026)
StatusOpen, Classified
ReasonBeckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and sit
Root causeSoftware design
ActionOn September 2, 2026, the firm began notifying customers via letters titled "URGENT MEDICAL DEVICE RECALL." Customers were instructed to inform their laboratory or medical director to assess laboratory and patient impact. For analyzers on software versions earlier than 1.5: - Users should verify that their instrument displays the intended units in the Test Menu page within the System Configuration page prior to reporting patient results. - Recall letter should be posted on or near affected analyzers if they have the affected software version. This issue is resolved in software version 1.5
Quantity146 US, 170 OUS
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K242190Access Cortisol; DxC 500i Clinical Analyzer
Beckman Coulter, Inc., Mar 5, 2025