Recall Z-0078-2027 ยท JAK System, X-Ray, Tomography, Computed

PHILIPS MEDICAL SYSTEMS

Spectral CT and Spectral CT 7500 RT; Software Version Numbers: 5.2.X (U.S.); 5.0.X, 5.2.X, 5.4.X (Outside U.S.); Model Numbers: 1. 728333 (U.S. and Outside U.S.), 2. 728340 (Outside U.S.), 3. 728344 (Outside U.S.), 4. 728345 (Outside U.S.);

DateOct 2, 2026 (initiated Aug 21, 2026)
StatusOpen, Classified
ReasonSoftware issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
Root causeUnder Investigation by firm
ActionOn August 21, 2026, URGENT Medical Device Correction letters were sent to customers. Customer actions: User Instructions Philips recommends the following actions be taken for continued safe use of the device: 1. Continue to use your system(s) in accordance with the intended use; 2. Circulate this notice to all users of the device so that they are aware of the product issues, associated potential hazards, and manufacturer recommendations (refer to Table 1 and Table 2 for specific details regarding recommended actions); 3. Post this notice near the affected unit(s) for ease of reference until a
Quantity539
DistributionWorldwide - US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, HI, IL, IN, KY, MA, MD, MN, NC, NY, OH, PA, TX, VA, WV and the countries of Australia, Austria, Belgium, Bulgaria, China, Croatia, Czech Republic, Denmark, Ecuador, El Salvador, France, Germany, Hong Kong, India, Indones
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K244008Spectral CT
Philips Medical Systems Technologies , Ltd., May 16, 2025