Recall Z-0306-2027 ยท ODG Endoscopic Ultrasound System, Gastroenterology-Urology

Pentax of America Inc

PENTAX Medical Ultrasound Endoscope. Model Number. EG36-J10UR.

DateOct 28, 2026 (initiated Sep 9, 2026)
StatusOpen, Classified
ReasonWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and during the procedure.
Root causeUnknown/Undetermined by firm
ActionAn Urgent Field Safety Notice dated 09/02/2026 was mailed to consignees. The notice informs consignees of two updates to the IFU of PENTAX Medical products, specifically the ultrasound endoscope devices and water bottle assembly. Consignees are to inform their staff of this notice and include the warning and updates to their IFUs. Additionally, consignees are instructed to change water for each procedure and fill the bottle approximately 2/3 full, or to the 200 marking on the bottle scale, with sterile water. If the water/balloon feeding function does not work consignees are to fill the water
Quantity84 units
DistributionWorldwide - US Nationwide and International distribution - Pending.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200090PENTAX Medical EG-J10U Endoscopic Ultrasound System - EG34-J10U Ultrasound Upper GI Video Scope (Convex Array Type), EG3
Pentax of America, Inc., Jul 23, 2020