Recall Z-3299-2026 ยท HIH Hysteroscope (And Accessories)

Olympus Corporation of the Americas

Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.

DateSep 25, 2026 (initiated Aug 24, 2026)
StatusOpen, Classified
ReasonThere are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
Root causeProcess control
ActionOlympus notified the single Japanese distributor on about 08/24/2026 via emailed "URGENT: FIELD SAFETY NOTICE." Consignees and end users are being instructed to store a copy of the provided correct sterilization parameters with current package insert, use only the corrected sterilization parameters, complete and return the provided reply form, and if units have been further distributed, notify those customers.
Quantity72 units
DistributionInternational distribution to the country of Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K890328ADULT RESECTOSCOPE
Circon Video, Feb 6, 1990