Recall Z-3241-2026 · LNH System, Nuclear Magnetic Resonance Imaging

Philips North America Llc

MR 5300 Model Number (REF): 1. 782110, 2. 782152;

DateSep 18, 2026 (initiated Aug 21, 2026)
StatusOpen, Classified
ReasonA software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Root causeSoftware design
ActionOn August 24, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: " To prevent this issue from occurring, o Use the standard 3D Vane scan protocol from the Philips Examcard Database, or o For custom 3D Vane scan protocols, use a maximum of 480 for the reconstruction matrix and use a maximum of 167 for the number of slices. " You may continue to use affected systems in accordance with their intended use. " Circulate this notice to all users or new owners of this device so that they are aware of the issues and associated hazard/harm. " Please retain this Ur
Quantity611
DistributionWorldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193215Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems
Philips Medical Systems Nederland B.V., Apr 10, 2020
K223442MR 5300 and MR 7700 R11 MR Systems
Philips Medical Systems Nederlands B.V., Dec 23, 2022
K251808Achieva; Intera; Ingenia 1.5T; Ingenia 3.0T; Ingenia 1.5T CX; Ingenia 3.0T CX; Ingenia Elition S; Ingenia Elition X; Ing
Philips Medical Systems Nederland B.V., Jul 11, 2025