ClearanceBrief
Recall Z-3298-2026 ยท OJF Hysterectomy Kit

Cardinal Health 200, LLC

Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT, LAVH , Catalog Number: PB63VHSTC

DateSep 25, 2026 (initiated Aug 18, 2026)
StatusOpen, Classified
ReasonThe kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.
Root causeUnder Investigation by firm
ActionAn URGENT: PRODUCT CORRECTION notification letter dated 8/18/26 was sent to customers. Presource has completed its investigation regarding this recall and determined that you may have received Presource kits containing the affected Stryker component(s) and lot number(s). Please reference the attached report identifying your facility s kits, along with the corresponding lot numbers that are affected by this action. If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits. 2. ALERT your clinicians of this action. 3. AFFIX a WARNING LABEL to the f
Quantity22 kits
DistributionUS Nationwide distribution in the states of CA, IL, KY, MO, NC, NY, OH, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.