Recall Z-0203-2027 ยท PHV Continuous Glucose Monitor Retrospective Data Analysis Software

Medtronic MiniMed, Inc.

CareLink Clinic, REF: MMT-7350

DateOct 21, 2026 (initiated Sep 11, 2026)
StatusOpen, Classified
ReasonCareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the pati
Root causeSoftware design
ActionOn 9/11/2026, correction notices were mailed to the health care professional, and customer asking them to do the following: The Customer was asked to do the following: Immediately contact your Healthcare Provider (HCP) to confirm your insulin settings values. Complete and return the confirmation form by visiting https://info.medtronicdiabetes.com/FA1587 The health care professional was asked to do the following: If you provided diabetes treatment recommendations to patients based on custom Total Daily Dose(TDD) or long-acting insulin values entered into the CareLin Clinic Settings Calculat
Quantity1
DistributionUS Nationwide distribution in the state of KS.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.