Recall Z-3251-2026 ยท OZD Temporary Non-Roller Type Left Heart Support Blood Pump
Abiomed, Inc.
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Control
| Date | Sep 21, 2026 (initiated Feb 19, 2019) |
| Status | Open, Classified |
| Reason | Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs. |
| Root cause | Under Investigation by firm |
| Action | No customer notifications were sent. The retrospectively reported recall does not currently affect any Automated Impella Controllers (AICs) in the field pending the correction. The hardware update was implemented to AICs in the field through servicing to address potential safety concerns. |
| Quantity | 3,322 units |
| Distribution | Worldwide - US Nationwide and the countries of Austria, Belgium, Canada, France, Germany, Hong Kong, Italy, Luxembourg, Poland, Spain, Switzerland, United Kingdom. |