ClearanceBrief
Recall Z-3283-2026 ยท KPS System, Tomography, Computed, Emission

Shanghai United Imaging Healthcare Co., Ltd.

Positron Emission Tomography and Computed Tomography System, Model: uMI 550 with software

DateSep 25, 2026 (initiated Aug 11, 2026)
StatusOpen, Classified
ReasonPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Root causeSoftware design
ActionOn 8/11/2026, correction notices were emailed to customers informing them that software will be released to resolve the software problem. Customers were asked to do the following: When using the Study Split function, follow the requirements below: 1) Temporarily disable the "Study Split" function. Navigate to the system configuration interface and uncheck the "Auto Study Split" option. 2) If the "Study Split" function is clinically necessary, open the system configuration interface and select "Split on PACS" instead of "Split on Local". 3) When "Auto study split" is checked, ensure that PACS (
Quantity34
DistributionUS Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX, ID, LA, MT, FL, CO, IN, KY, AL, OH, MI, IA, MO, NY, NV, TN, WV, SC, WI, IL, MS, MA, NJ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K182237uMI 550 PET/CT System
Shanghai United Imaging Healthcare Co., Ltd., Oct 12, 2018
K193241Emission Computed Tomography System
Shanghai United Imaging Healthcare Co., Ltd., Jan 26, 2020