Recall Z-3282-2026 ยท JAK System, X-Ray, Tomography, Computed

Shanghai United Imaging Healthcare Co., Ltd.

Computed Tomography X-ray System, Model: uCT ATLAS Astound with software

DateSep 25, 2026 (initiated Aug 11, 2026)
StatusOpen, Classified
ReasonPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Root causeSoftware design
ActionOn 8/11/2026, correction notices were emailed to customers informing them that software will be released to resolve the software problem. Customers were asked to do the following: When using the Study Split function, follow the requirements below: 1) Temporarily disable the "Study Split" function. Navigate to the system configuration interface and uncheck the "Auto Study Split" option. 2) If the "Study Split" function is clinically necessary, open the system configuration interface and select "Split on PACS" instead of "Split on Local". 3) When "Auto study split" is checked, ensure that PACS (
Quantity53
DistributionUS Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX, ID, LA, MT, FL, CO, IN, KY, AL, OH, MI, IA, MO, NY, NV, TN, WV, SC, WI, IL, MS, MA, NJ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203448uCT ATLAS with uWS-CT-Dual Energy Analysis
Shanghai United Imaging Healthcare Co., Ltd., Aug 11, 2021
K231482uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual Energy Analysis
Shanghai United Imaging Healthcare Co., Ltd., Nov 29, 2023
K243617uCT ATLAS Astound with uWS-CT-Dual Energy Analysis; uCT ATLAS with uWS-CT-Dual Energy Analysis
Shanghai United Imaging Healthcare Co., Ltd., May 16, 2025