Recall Z-3249-2026 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Philips Ultrasound, LLC

Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234

DateSep 18, 2026 (initiated Aug 7, 2026)
StatusOpen, Classified
ReasonDue to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
Root causeSoftware Design Change
ActionOn August 7, 2026, Philips issued a Urgent Medical Device Correction Notification to affected consignees. Philips requested consignees take the following actions: 1. Due to possibility of a weak PW Doppler Signal on EPIQ Elite running 14.0 SW, customers are advised not to use C5-1 Abd Vasc Tissue Specific Preset (TSP). Please use a different TSP (e.g., Abd Gen, Abd Renal) within C5-1. Please note that this issue (i.e., Weak PW Doppler signal) is restricted to the Abd Vasc TSP with C5-1 on EPIQ Elite running SW 14.0 only. 2. Otherwise, you may continue to use your device in accordance with th
Quantity219 units
DistributionWorldwide - US Nationwide distribution in the states of AL, CA, CT, FL, GA, KS, MD, MN, NJ, NY, OH, PA, SC, TN, TX, VA, WA, WI and the countries of Austria, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lithuania, Netherlands, Norway, Philippines, Poland, Romania, Slovenia, South Korea,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K260123EPIQ Series Diagnostic Ultrasound System
Philips Ultrasound, LLC, May 27, 2026
K260680EPIQ Series Diagnostic Ultrasound Systems; Affiniti Series Diagnostic Ultrasound Systems
Philips Ultrasound, LLC, May 27, 2026