Recall Z-3321-2026 ยท NHJ Device, Positive Pressure Breathing, Intermittent
Percussionaire Corporation
Intrapulmonary Percussive Ventilation, IPV-1C, Product Number: S00002-C; Phasitron 5, Product Number: P5 used with Instruction Manual
| Date | Sep 28, 2026 (initiated Aug 18, 2026) |
| Status | Open, Classified |
| Reason | Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation |
| Root cause | Device Design |
| Action | On 8/18/2026, correction notices were emailed to customers and distributors who were informed of the following: The following warning will be added to associated Instructions for Use: Warning: Do not use IPV therapy in conjunction with volatile anesthetic delivery systems. The high-frequency percussive flow generated during IPV therapy may alter the controlled delivery of volatile anesthetic agents within the breathing circuit, resulting in unpredictable anesthetic agent concentrations that may result in serious patient harm. Note: Consult the anesthesia healthcare provider before disconnect |
| Quantity | 229,105 |
| Distribution | Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, C |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.