ClearanceBrief
Recall Z-0075-2027 ยท FRC Indicator, Biological Sterilization Process

Steris Corporation

VERIFY Biological Steam Test Pack Model/Catalog Number: S3069

DateOct 2, 2026 (initiated Aug 31, 2026)
StatusOpen, Classified
ReasonGeobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Root causeUnder Investigation by firm
ActionFirm began notifying customers on August 31, 2026 via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to destroy any affected product on hand. The firm will replace any product still within its shelf life.
Quantity233
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K854902SURGICOT STEAM TEST PACK
Surgicot, Inc., May 20, 1986