ClearanceBrief
Recall Z-0067-2027 ยท OGE Epidural Anesthesia Kit

Custom Interventional Pain Management, LLC

Brand Name: Ferndale Laboratories; Medline; Custom Interventional Pain Management CIPM Product Name: Mastisol Skin Adhesive Model/Catalog Number: 24161B (Ferndale and Medline); Tray 17025 17g continuous PBN tray (CIPM) Software Version: n/a Product Description: Skin Adhesive in a narrow plastic sterilized tube. Mastisol is a skin glue often used during surgical procedures. Continuous epidural tray

DateOct 1, 2026 (initiated Jan 15, 2026)
StatusOpen, Classified
ReasonDue to manufacturing issues with the liquid skin adhesive containing in surgical convenience trays/kits.
Root causeProcess control
ActionOn 01/15/2026, a customer notification letter was email to a single customer informing them that, the MASTISOL Liquid adhesive shipped in kits in lot 1225271, P.O.#2513972 was exposed to a double sterilization cycle by Ferndale Laboratories, Inc. which may contribute to the butyrate tubing being brittle and cracking during actuation. The customer was instructed to: 1. lmmediately examine your inventory and quarantine Lot 1225271. 2. Return all units of MASTISOL in Lot 1225271 to: CIPM 12460 Network Blvd. Suite 106 San Antonio, TX 78249 3. Complete the enclosed Return Materials Authorization F
Quantity56
DistributionU.S. Nationwide distribution in the state of TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.