ClearanceBrief
Recall Z-0007-2027 ยท KCY Tourniquet, Pneumatic

ASP Global, LLC.

Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 15" X 4" 2-LINE (CS 10) GRAY Model/Catalog Number: 564005 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in sur

DateOct 1, 2026 (initiated Aug 24, 2026)
StatusOpen, Classified
ReasonDue to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Root causeUnder Investigation by firm
ActionOn 08/24/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was emailed to customers informing them that, ASP Global identified a discrepancy between the labeled shelf life and available shelf-life validation data for certain Sterile Disposable Tourniquet Cuff products. While affected products were labeled with a 10-year shelf life, available validation documentation supports a shelf life of 5-years. Customer are instructed to: a) Review Appendix A to identify affected item codes and lot numbers. b) Discontinue use of the affected products. c) Quarantine any remaining impacted inventory to preven
Quantity70
DistributionUS Nationwide distribution in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.