Recall Z-3149-2026 ยท OZD Temporary Non-Roller Type Left Heart Support Blood Pump

Abiomed, Inc.

Impella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella Optical Controller, Packaged, AU; Product Code: 0042-0010-AU. 2. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 3. Impella Optical Controller, Packaged, IN; Product Code: 0042-0010-IN. 4. Impella Optical Controller, Packaged, UK; Product Code: 0042-0010-UK. 5. Impella O

DateSep 16, 2026 (initiated Aug 12, 2026)
StatusOpen, Classified
ReasonThe potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
Root causeUnder Investigation by firm
ActionOn August 12, 2026 URGENT MEDICAL DEVICE RECALL (REMOVAL) letters were sent to customers. Actions to be Taken: 1. Hospital inventory can continue to be used while awaiting service from Abiomed to ensure continuity of care. 2. Abiomed s field service team will contact you to schedule the specified replacement. 3. In the event of these purge flag component failures and associated alarms, as per IFU, first reinsert purge disc to address the alarm. If that action does not clear the alarm, switch to a backup Automated Impella Controller. 4. Ensure cleaning of the AIC housing with mild detergent a
Quantity1422 units (487 US, 935 OUS)
DistributionDomestic: Nationwide Distribution. International: United Arab Emirates, Austria, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czechia, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Ir
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.